Coating Thickness Gauge Leeb222

  • High quality metal probes
  • Two measuring methods:continuous and single
  • Two working mode: direct and batch
  • Limit setting function
  • Switch off automatically or manually

For product pricing, customization, or other inquiries:

Share to

                                                     Functions & Features                                              

  • High quality metal probes
  • Two measuring methods:continuous and single
  • Two working mode: direct and batch
  • Limit setting function
  • Switch off automatically or manually

                                                      Measuring Materials                                               

Magnetic Induction (Fe): Measuring the thickness of Non-magnetic coating on magnetic metal substrate, such as aluminum, chromium, copper, zinc, rubber, paint on the base of steel, iron, alloy and magnetic steel

. Eddy Current (NFe): Measuring the thickness of Non-conductive coating on non-magnetic metal substrate, such as rubber, plastic, paint, oxide on the base of aluminum, copper, zinc, tin.

ModelLeeb210Leeb211Leeb220Leeb221Leeb222
Measuring principleFeNFeFeNFeFe & NFe
Measuring Range0~1250μm
ProbeSettledChangeable
ShellPlastic
Accuracy±[(1~3%)H+1] μm H refers to the thickness of testing piece
Minimum resolution (µm)0.1μm
Min curvature of the min area (mm)Convex1.5  Concave9
Diameter of the min area (mm)Φ7
Critical thickness of substrate (mm)0.5
Memory200 groups measured data
Dimensions115×70×30mm
Power supplyAAA  Alkaline battery
  Standard Configuration  Main Machine,5 calibration specimens (48.5μm、99.8μm、249μm、513μm、1024μm),1 probe & substrate ( 2 probes & substrates for Leeb222)
Optional AccessoriesProbes,Specimens